Dwyer 400-23-GAGE

Dwyer 400-23-GAGE gaisa ātruma mērītāja lietošanas instrukcija

Model: 400-23-GAGE

Ievads

This manual provides essential instructions for the proper setup, operation, and maintenance of your Dwyer 400-23-GAGE Air Velocity Meter. Please read this manual thoroughly before using the device to ensure accurate measurements and safe operation. Retain this manual for future reference.

Produkts beidziesview

The Dwyer 400-23-GAGE is a precision instrument designed for measuring air velocity. It typically includes a manometer, pitot tube, tubing, and necessary accessories for field use. This device is suitable for various industrial and scientific applications requiring accurate airflow measurement.

Dwyer 400-23-GAGE Air Velocity Meter kit including manometer, pitot tube, tubing, and carrying case.

1. attēls: Dwyer 400-23-GAGE Air Velocity Meter kit. This image displays the complete kit, including the inclined manometer, a pitot tube, connecting tubing, and a protective carrying case. Also visible are small bottles, likely containing manometer fluid, and instruction sheets.

Iestatīšana

  1. Izpakošana: Uzmanīgi izņemiet visas sastāvdaļas no iepakojuma. Pārliecinieties, vai visas iepakojuma sarakstā uzskaitītās preces ir klāt un nav bojātas.
  2. Manometer Fluid Filling: If the manometer is not pre-filled, carefully add the specified manometer fluid (typically included) to the reservoir until it reaches the indicated zero mark. Refer to the specific instructions on the manometer itself for precise filling procedures. Avoid overfilling or introducing air bubbles.
  3. Caurules savienojums: Connect one end of the flexible tubing to the high-pressure port of the manometer and the other end to the total pressure port of the pitot tube. Connect a second piece of tubing from the low-pressure port of the manometer to the static pressure port of the pitot tube. Ensure all connections are secure and airtight.
  4. Pitot Tube Placement: Position the pitot tube in the airflow stream at the desired measurement point. Ensure the tip of the pitot tube is facing directly into the airflow for accurate readings.
  5. Leveling the Manometer: Place the manometer on a stable, level surface or mount it securely. Use the integrated leveling bubble (if present) or a separate level to ensure the manometer is perfectly horizontal. This is critical for accurate inclined manometer readings.

Lietošanas instrukcijas

  1. Nulles regulēšana: Before taking any measurements, ensure the manometer fluid level is at the zero mark when no pressure is applied. Adjust the zeroing knob if necessary.
  2. Lasījumu pieņemšana: With the pitot tube correctly positioned in the airflow, observe the fluid level in the manometer. The difference in fluid height indicates the differential pressure.
  3. Velocity Calculation: Use the provided air velocity calculation chart or formula (typically printed on the manometer or in accompanying documentation) to convert the differential pressure reading (in inches of water) into air velocity (e.g., feet per minute or meters per second). This calculation often requires knowing the air density, which can vary with temperature and atmospheric pressure.
  4. Vairāki lasījumi: For improved accuracy, take several readings at different points within the airflow duct or area and average them.
  5. Vides apsvērumi: Ensure the operating environment is within the specified temperature and humidity ranges for the instrument to maintain accuracy.

Apkope

  • Šķidruma līmeņa pārbaude: Regularly check the manometer fluid level and replenish if necessary. Use only the manufacturer-specified fluid.
  • Tīrīšana: Keep the manometer and pitot tube clean. Use a soft, damp cloth for the exterior. Ensure the pitot tube openings are free from obstructions.
  • Cauruļu pārbaude: Periodically inspect the connecting tubing for cracks, kinks, or leaks. Replace damaged tubing immediately to prevent inaccurate readings.
  • Uzglabāšana: When not in use, store the instrument in its protective carrying case in a clean, dry environment, away from extreme temperatures and direct sunlight. Ensure the manometer fluid is secured to prevent spillage.
  • Kalibrēšana: For critical applications, periodic calibration by a qualified service center is recommended to maintain measurement accuracy.

Problēmu novēršana

  • Nav rādījumu vai rādījumi ir neregulāri:
    • Check for kinks or blockages in the tubing.
    • Ensure tubing connections are secure and airtight.
    • Verify the pitot tube is correctly positioned in the airflow.
    • Check manometer fluid level and for air bubbles.
    • Ensure the manometer is level.
  • Neprecīzi rādījumi:
    • Confirm the manometer is properly zeroed.
    • Verify the correct air density is used in velocity calculations.
    • Check for any physical damage to the pitot tube or manometer.
    • Consider professional calibration if accuracy issues persist.

Specifikācijas

Modelis400-23-GAGE
ZīmolsDvaiers
RažotājsDvaiers
Iepakojuma izmēri13 x 3 x 3 collas
Iepakojuma svars17.4 mārciņas
ASINB077NJWHYB

Note: Specific measurement ranges and accuracy details are typically found on the device itself or in supplementary documentation.

Saistītie dokumenti — 400-23-GAGE

Iepriekšview Dwyer sērijas 477AV rokas digitālais manometrs: specifikācijas, uzstādīšana un lietošanas instrukcijas
Visaptverošs Dwyer sērijas 477AV rokas digitālā manometra ceļvedis, kurā ietvertas specifikācijas, uzstādīšana, lietošanas instrukcijas, funkcijas un apkope spiediena, plūsmas un ātruma mērījumiem.
Iepriekšview Dwyer 160G sērijas gaisa ātruma režģa un 485B sērijas rokas temperatūras/mitruma mērītāja specifikācijas
Detalizētas specifikācijas Dwyer 160G sērijas vidējā gaisa ātruma režģim un 485B sērijas rokas digitālajam temperatūras un mitruma mērītājam, tostarp funkcijas, pielietojums un tehniskie parametri.
Iepriekšview Dwyer Magnehelic diferenciālā spiediena mērītājs — biļetens A-27
Dwyer Magnehelic diferenciālā spiediena mērītāju, 2000. sērijas, tehniskās specifikācijas, uzstādīšanas rokasgrāmata, darbība un problēmu novēršana.
Iepriekšview Dwyer Magnehelic diferenciālā spiediena mērītāja biļetens A-27
Tehniskais biļetens, kurā sīki aprakstīts Dwyer Magnehelic diferenciālā spiediena mērītājs, tostarp specifikācijas, uzstādīšana, darbība un problēmu novēršana dažādiem modeļiem.
Iepriekšview Dwyer sērijas A3000 fotospirālveida diferenciālā spiediena slēdža/mērītāja lietotāja rokasgrāmata
Dwyer sērijas A3000 fotospirālveida diferenciālā spiediena slēdža/mērītāja lietotāja rokasgrāmata, kurā sīki aprakstītas tā specifikācijas, uzstādīšana, elektriskie savienojumi un apkope.
Iepriekšview Dwyer sērijas TUF ultraskaņas enerģijas skaitītāji: plūsmas un temperatūras monitorings
Detalizētas specifikācijas, funkcijas, pielietojumi un pasūtīšanas informācija Dwyer sērijas TUF ultraskaņas enerģijas skaitītājiem, kas paredzēti precīzai plūsmas un temperatūras uzraudzībai ar Modbus® vai BACnet komunikāciju.